Scanning electron microscope testing

For dimensional work on a scanning electron microscope, the scale assigned to the stored image matters as much as the magnification label. Identify the image acquisition settings and the range of feature sizes routinely measured by the laboratory.

Send an enquiry
Scanning electron microscope — equipment illustration
Equipment illustration
On this page

Checking image scale in the scan directions

For an image-scale comparison, send a reference image in its original format, with the scan dimensions and pixel size. A resized image copied into a report can lose the relationship between pixels and length. The pitch artefact’s certificate, the scan direction and the imaging settings let us check the displayed scale at the feature sizes your analysts measure.

Scanning electron microscope dimensional scale

The result records magnification, working distance and the operating settings used by the laboratory’s method. Image-scale verification addresses the dimensions calculated from the image; elemental analysis has its own reference and method requirements. Nominal magnification is retained as a setup setting alongside the measured scale.

Specifications and service checks

If measurements are taken after resizing an exported image, the on-screen magnification label alone cannot reproduce the scale. Keep the original image dimensions and the acquisition settings alongside the reference image. This allows the dimensional scale in the measurement software to be checked.

Range and configuration
JEOL lists spatial resolution separately from magnification on the JCM-7000. For a dimensional image check, supply the field width and measurement scale used by the operator.
Manufacturers and models
The Evident CX23 is an optical microscope, while the JEOL JCM-7000 NeoScope is a benchtop scanning electron microscope.
What we check
Compare the image scale against a dimensional reference at the settings where measurements are made. Record whether the result applies to the eyepiece, camera or saved image.
Reference equipment and accuracy
Choose a certified pitch or dimensional artefact suitable for the SEM field of view and feature sizes. Its assigned geometry and uncertainty support an image-scale comparison.
When to repeat the check
Recheck after changing optics, camera settings or image calibration software. Keep the scale results for each configuration used for dimensional work.

Technical references Evident CX23 microscope instructions · JEOL instrument catalogue