Semi-micro balance calibration

A semi-micro balance may provide its finest division only over part of its capacity. A tared weighing vessel still contributes to the gross load and can push the instrument into a coarser range. Check the actual division displayed with the vessel in place.

Send an enquiry
Semi-micro balance — equipment illustration
Equipment illustration
On this page

Fine and coarse weighing ranges

Provide the fine and coarse range information if marked, together with the smallest net quantity the procedure uses. If the working load sits close to a range transition, include readings either side. The report should identify which division applied at each requested point.

Some semi-micro balances display an extra decimal only over part of their capacity. If routine work uses a heavy tare container, note the display step with that container in place. A check planned solely from the light-load specification can overlook the range used at the bench.

Specifications and service checks

A sample can fit comfortably within the capacity yet use a coarser readability when its heavy container is present. Record the loaded pan value and the display increment at that load before selecting a test for small ingredient additions.

Range and configuration
OHAUS’s EXR125D has two capacity bands, 82 g and 120 g, with respective readabilities of 0.01 mg and 0.1 mg. Identify which band contains the tared sample.
Manufacturers and models
OHAUS Explorer EXR125D is a dual-range example. The larger capacity on the plate should not be quoted as the capacity at the finer division.
What we check
Check repeatability and error in each band that the laboratory uses. Make the transition between bands identifiable in the results.
Reference equipment and accuracy
Use certified low-uncertainty weights appropriate to each display interval and include the uncertainty at the smallest requested net mass.
When to repeat the check
Track low-range performance separately when reviewing an annual recall. Relocation or damage to the draft shield warrants an interim check.

Technical references OHAUS Semi micro balance reference