Identifying individual pipettes within a lot
If the pipettes are supplied by lot, define how many pieces are to be measured and how each result will be identified. Report the graduation interval checked and the specified delivery convention. This prevents an ordinary drain result being compared with a result obtained using blow-out.
A comparison made by draining to the tip and one made by following a marked blow-out instruction can yield different delivered volumes. If the laboratory changed supplier since the last lot check, read the new inscription before reusing the earlier sampling sheet or delivery instruction.
Specifications and service checks
Some serological pipettes have scale markings extending towards the tip and a marked blow-out convention. Copy the scale direction and delivery marking from the actual item. A volume read from the upper graduation to the tip depends on how that pipette is meant to drain.
- Range and configuration
- Corning catalogue number 4488 is a 10 mL Stripette serological pipette. Note the scale direction, lot and stated delivery convention for the version in use.
- Manufacturers and models
- Corning’s Stripette is a disposable graduated serological pipette. Retain the lot and size on the packet when requesting a delivery check.
- What we check
- Measure the amount delivered under the drainage convention marked on the instrument. Record which marks were used and whether blow-out is specified for that design.
- Reference equipment and accuracy
- A traceably calibrated balance and thermometer support a volume comparison. Choose resolution and evaporation control for the smallest delivery under test.
- When to repeat the check
- Identify glass pieces individually where practical. For disposable serological or transfer pipettes, follow the lot-control requirements used by the laboratory.
Technical references BRAND volumetric glassware testing guide · Corning Stripette serological pipettes · NIST volumetric calibration procedures · Corning 4488 10 mL Stripette