On this page
Setting depth and ring orientation
Mark the usual measuring depth and any orientation line on the ring. State whether results at multiple heights or directions are needed. Supply the nominal diameter and the uncertainty needed for the instrument being set, along with any previous values used in the shop.
If the ring lives beside two bore gauges of different reach, one head may seat deeper than the other. Identifying both working bands lets the laboratory associate the result with the position used by each gauge.
Specifications and service checks
If the assigned working bore is 40.002 mm, a head preset to 40.000 mm in that ring acquires a 0.002 mm offset relative to the assigned value. Decide how the shop enters the ring value before comparing two head settings.
- Range and configuration
- A ring has one assigned internal diameter. A setting ring establishes a numerical bore-gauge setting while a limit ring carries a GO or NO-GO function for shafts.
- Manufacturers and models
- Vermont Gage supplies cylindrical rings and Mahr supplies setting gauges. A threaded ring has additional features and requires its thread designation.
- What we check
- Determine the actual inside diameter in the plane used to set the gauge and record the temperature basis. Measurements at further planes and orientations show whether the setting changes with contact position.
- Reference equipment and accuracy
- Use a calibrated internal length machine or comparator with suitable setting standards. Choose the reference uncertainty from the ring tolerance or the uncertainty required of the downstream bore measurement.
- When to repeat the check
- Check for raised edges and handling damage. Re-establish the internal diameter after lapping or if successive bore heads no longer agree on a known part.
Technical references Vermont Gage Plain cylindrical gauges · Mahr Plain ring gauge reference