Single channel pipette calibration

A single channel pipette has one delivery path, so results can be followed directly against the body and fitted tip. For an adjustable model, performance at the lower working setting may be more important to the technician than performance at full capacity.

Send an enquiry
Single channel pipette — equipment illustration
Equipment illustration
On this page

Mean delivery and repeatability at the working volume

List the volumes actually selected on the bench and the criterion for each. Keep repeated delivery results distinguishable from their mean. Seeing both helps a user tell an inconsistent delivery from a stable difference that may call for adjustment.

For a nominal 100 µL transfer, repeated results clustered near 98 µL suggest a different issue from readings scattered either side of 100 µL with an average close to target. Ask for the mean and repeatability at the working volume. They let the person reviewing the certificate see the difference before deciding what to do with the pipette.

Specifications and service checks

When one pipette is shared between reagent preparation and small aliquots, each user may care about a different setting. Gather both working volumes before booking. This lets the same handle be checked at the settings people actually dial rather than only at its upper marking.

Range and configuration
The Gilson P200 M example spans 20–200 µL. Record both volume limits and the tip specified for the particular pipette. One increment on the volume display is not its allowable delivery error.
Manufacturers and models
Gilson PIPETMAN Classic and PIPETMAN M are examples of mechanical and electronic families. The complete model code identifies the volume range.
What we check
Check repeated deliveries from the one identified outlet at the selected settings. Report the mean delivered volume and the variation between deliveries so the operator can see which changed.
Reference equipment and accuracy
Choose a calibrated balance with resolution and uncertainty suited to the smallest delivery. Account for water temperature, evaporation and air buoyancy when converting mass to volume.
When to repeat the check
Use the frequency of use, local checks and previous results to set recall. Give a leaking or repaired pipette a fresh check before returning it to the bench.

Technical references Gilson PIPETMAN Classic guide · Gilson PIPETMAN M specifications · ISO 8655-6 gravimetric volume determination