Smart pressure transmitter calibration

A smart pressure transmitter may supply both a digital process value and an analogue output. The digital indication can agree with the applied pressure while the analogue trim or configured output range still needs attention.

Send an enquiry
Smart pressure transmitter — equipment illustration
Equipment illustration
On this page

Digital process value and analogue loop current

Include the tag, lower and upper range values, damping setting and which output the plant uses. If readings are taken through a handheld communicator, distinguish that digital reading from the current measured at the transmitter terminals.

If the transmitter’s digital value agrees with reference pressure but the control room shows a different value, retain the simultaneous measured loop current. It helps distinguish an analogue output discrepancy from an error in the receiving input or its scaling.

Specifications and service checks

A maintenance laptop can read a digital process variable while the panel reads an analogue current. Note both values at the same stable condition and identify the assigned primary variable. This short comparison is especially useful after a configuration download or range change.

Range and configuration
Specify the sensor limits and the configured lower and upper values separately. On a linear 4–20 mA range, the mid-span output is 12 mA, even when zero pressure is outside the configured span.
Manufacturers and models
Emerson's Rosemount 3051 family includes differential, gauge and absolute versions with several output options. Use the complete model code.
What we check
Read the digital process value as well as loop current. Compare both with the applied pressure and retain the stored lower and upper range values, damping and any square-root setting.
Reference equipment and accuracy
Use a calibrated pressure reference of the correct mode. A calibrated current meter is also needed for a loop-output check. Both uncertainties enter the result.
When to repeat the check
Review the interval against drift in the working range. Check again after a sensor change, rerange or repair when it affects the reported value.

Technical references Product Data Sheet: Rosemount 3051 Pressure Transmitter