Soldering bath calibration

A soldering bath is usually controlled at one point while production work takes place over a larger surface. If the accepted temperature applies to a particular working position, mark that location and specify the depth at which it is defined.

Send an enquiry
Soldering bath — equipment illustration
Equipment illustration
On this page

Temperature at the normal working depth

If operators use the left end for one component and the right end for another, those two positions may warrant separate results even though the bath has one controller. Mark the distance from an edge and the working depth for each. Include the bath material and usual hold temperature. This gives the next inspection a position it can actually return to.

Soldering baths

State whether the requirement concerns the controller indication or a comparison within the working bath at named positions. Surface condition and access can affect how a reference is placed. RTS compares temperature at the specified bath positions and load condition, recording the setpoint and measured values separately.

Specifications and service checks

A thermometer placed beside the pot wall can read differently from one at the normal immersion point of the work. Mark the point and depth used for the process check, along with the working setpoint. This makes a repeat visit comparable after the pot has been reinstalled.

Range and configuration
Use the working temperature of the solder alloy and the controller’s rated span. Heater power, pot volume and melting point describe different properties of the equipment.
Manufacturers and models
JBC’s CTS stand accepts T245 and T470 handles fitted with solder pot cartridges. Identify the cartridge, handle and controller on the service request; the stand designation alone gives no bath temperature.
What we check
Compare the bath temperature at a defined position and load with a compatible reference probe. Record stability during the agreed measurement period.
Reference equipment and accuracy
A calibrated sheathed temperature probe and readout suited to the alloy and temperature are needed. A surface IR measurement has additional emissivity and oxide-film effects.
When to repeat the check
Deposits, heater changes or a replacement sensor warrant a fresh temperature check. Review the recall against the allowed process window.

Technical references JBC solder pot stand manual · BIPM guide to industrial platinum resistance thermometers