A displaced line or a change in signal level?
If a known spectral line appears displaced, send its assigned and observed positions with the units. If the line is in the right place but its signal is low, retain the acquisition settings from that run. A change in integration time can change the recorded intensity. For a reported concentration, tell us the material and analytical method so we can identify the references relevant to that result.
Identify the spectrometer technique
Give the technique, model, detector and reported quantity, together with the method used by the laboratory. Wavelength-related checks, response checks and a material-composition method are different requests. RTS checks the named functions using the reference appropriate to that optical or analytical technique. Include relevant reference-material certificates and avoid describing a general instrument service as validation of every stored application method.
Specifications and service checks
Two features can appear at the correct wavelengths while their relative heights change after a fibre is replaced. Identify whether the issue concerns line position, relative intensity or absolute optical response. Each calls for a different comparison.
- Range and configuration
- The Ocean Optics Flame guide lists 200–1025 nm for its No. 31 UV–NIR grating. The instrument serial number and installed grating establish which lines fall on a particular detector.
- Manufacturers and models
- Ocean Optics Flame configurations cover different bands. Keep the spectrometer, grating and fibre identifiers with any certificate intended to cover spectral intensity.
- What we check
- For wavelength calibration, compare the displayed positions of identifiable emission lines with their reference wavelengths. A request for absolute irradiance also covers the response of the complete fibre, probe and detector assembly.
- Reference equipment and accuracy
- Choose a line source with well-characterised wavelengths in the instrument’s band. Intensity work requires a source with calibrated spectral irradiance and the optical geometry used for the result.
- When to repeat the check
- Monitor a reference peak between calibrations. Reassess wavelength assignments after optical realignment, and repeat radiometric calibration after a fibre or collection optic is changed.
Technical references Ocean Optics Spectrometer reference · Ocean Optics Calibration FAQs · Ocean Optics Spectrometer Light Sources