Spectrophotometer testing

A spectrophotometer can be checked for wavelength indication and for photometric response. A known absorption feature locates a wavelength, while a photometric reference tests a value such as absorbance or transmittance. Those checks address different parts of an analytical method.

Send an enquiry
Spectrophotometer — equipment illustration
Equipment illustration
On this page

Comparing wavelength and absorbance

If the wavelength of a peak has moved, a wavelength reference helps locate the shift. If the position is steady but absorbance has changed, a photometric reference at the working wavelength is more informative. Give the cell path length and the holders used. A laboratory that always reads a blank in position one and a sample in position two may need both positions identified on the report.

Spectrophotometer functions to specify

Identify the cell or sample holder, path length where applicable and the references specified for each check. Wavelength performance and photometric response are separate characteristics. Add any method-specific qualification requirement explicitly instead of assuming a single reference reading covers the complete instrument.

Specifications and service checks

A filter with 10% transmittance corresponds to an absorbance of 1 under the usual base-ten definition. Use the assigned value and wavelength of the reference when comparing functions. A wavelength check alone cannot establish the displayed absorbance.

Range and configuration
Shimadzu lists 190–1100 nm and 1 nm resolution for the UV-1900i Plus. Wavelength is reported in nanometres, while the photometric result can be absorbance or percentage transmittance.
Manufacturers and models
The Shimadzu UV-1900i Plus is a double-beam UV–visible instrument. A model with an integrating sphere or another accessory may need an additional configuration recorded.
What we check
Check wavelength indication with certified spectral features and the photometric scale with certified absorbance or transmittance filters. State which wavelengths, bandwidths and optical paths were used. Stray light and bandwidth require their own tests if requested.
Reference equipment and accuracy
The wavelength reference needs assigned peak positions. Select photometric filters or solutions with certified values at the actual measurement wavelengths and uncertainties appropriate to the acceptance limits.
When to repeat the check
Check a reference material at wavelengths used in routine work. Review the calibration after a lamp or detector change, or if a control filter develops a consistent shift even with fresh blanks.

Technical references Shimadzu UV-visible spectrophotometry · Shimadzu UV-visible spectrophotometry