Spectroscope calibration

A spectroscope may show lines for visual identification or provide a wavelength scale that can be read numerically. Confirming the location of known spectral lines is different from measuring the intensity of emitted light.

Send an enquiry
Spectroscope — equipment illustration
Equipment illustration
On this page

Line position and spectral resolution

Describe the optical input, scale or detector and the wavelength region used. If resolving two close lines is important, identify that requirement separately from the wavelength indication. For a visual instrument, include the slit or grating configuration used in practice.

A spectroscope used to distinguish two nearby lines can show the correct wavelength for an isolated line yet leave the close pair unresolved. When separation is the issue, give the two reference wavelengths and the slit or grating normally used. That is a more useful request than a single nominal wavelength.

Spectroscopes with a wavelength scale

Identify the wavelength features used in your work, any reference lamp and the viewing or input optics. A request to identify spectral lines differs from a request for radiometric amplitude results. State whether the required result is wavelength, radiometric response or both so the report describes the measurement clearly.

Specifications and service checks

When a student or inspector reads a line against a scale, the width of the line can determine how repeatably its centre is located. Identify the slit setting and spectral region for any complaint about line position, particularly when adjacent lines partly overlap.

Range and configuration
Quote the scale increment and wavelengths of interest. The ability to see two lines separately is a different characteristic from assigning a wavelength to either line.
Manufacturers and models
A. Krüss HS1504 is a hand spectroscope with a 400–750 nm scale marked in 10 nm divisions. Some viewing spectroscopes have no scale.
What we check
Compare identified spectral lines at several places on the scale. For a visual instrument record the slit setting and spread of repeated readings.
Reference equipment and accuracy
Use a suitable line source with recognised reference wavelengths. Scale reading and line width set practical limits to uncertainty.
When to repeat the check
Repeat the wavelength check after adjustment of the slit, prism or scale.

Technical references A. Krüss HS1504 spectroscope

Enquire about Spectroscope