Step spacing and probe direction
List the spacings or heights used for verification and the probe approach direction. If the device checks a coordinate measuring machine, include its normal support and orientation. A table of coordinates from the datum is different from a table of distances between neighbouring faces.
If a machine checks its travel at several shoulders along a step gauge, send the list of faces used and the designated origin. An error in one interval and an error in overall length answer different questions when an inspector compares the machine with the reference.
Specifications and service checks
One programme may read all step faces from the first face, while another measures only adjacent intervals. If a position discrepancy appears far down the travel, specify whether the required reference is cumulative distance or the single interval at that position.
- Range and configuration
- Give the assigned distance between selected step faces, their order and measurement direction. The physical length of the rail can exceed its calibrated span.
- Manufacturers and models
- Specify the maker and step layout on the gauge certificate. A machine-axis step standard and a single stepped shop gauge need different setups.
- What we check
- Compare selected faces on a length measuring machine or by interferometric displacement. Report each interval from the agreed datum and identify reversals.
- Reference equipment and accuracy
- Use traceable length standards or a laser reference. Thermal expansion, contact direction and the support positions matter on long multi-step artefacts.
- When to repeat the check
- Use the same supports during intermediate checks. Reassess after a knock, repair to an insert or a change in the datum feature.