Suction apparatus testing

A suction unit has a vacuum indication and a flow path through its jar, filter and tubing. The maximum indicated vacuum and the flow obtained through an assembled path are different performance measurements.

Send an enquiry
Suction apparatus — equipment illustration
Equipment illustration
On this page

Vacuum indication and flow through the assembly

A gauge reading in kPa and a flow reading in litres per minute belong on separate lines. For an engineering check of the assembled unit, tell us which jar, filter, tubing and regulator are fitted. If an earlier job covered only the gauge, mark that on the repeat order. A correct gauge indication alone cannot explain a change in flow after a filter was replaced.

Suction apparatus pressure and flow enquiries

Distinguish a vacuum indication from a flow measurement and from a complete functional assessment. Include the manufacturer’s test conditions and the department’s preparation declaration. The quotation should state the accepted quantities and configuration so the biomedical reviewer can see exactly what the resulting report covers.

Specifications and service checks

A dial on the pump body and a gauge at the collection line can show different values when flow is present. Identify the measuring point and the fitted tubing if the engineering enquiry concerns that difference.

Range and configuration
Laerdal’s service data lists a vacuum magnitude of 80 to 500+ mmHg for the LSU. Vacuum is measured relative to ambient pressure. Specify the units and sign convention plus any separate free-flow requirement on the service worksheet.
Manufacturers and models
Laerdal’s LSU is a portable suction unit with selectable vacuum indications.
What we check
Compare the vacuum indication to an independent pressure reference using the authorised service arrangement. Measure flow separately if that is a requested performance characteristic.
Reference equipment and accuracy
Use a calibrated vacuum gauge over the magnitude being tested. A suitable flow standard is needed for a flow result, with the accessories and test condition identified.
When to repeat the check
Repeat performance checks after pump, valve or gauge work. The facility’s service plan also covers battery and mechanical condition.

Technical references Laerdal product reference

Enquire about Suction apparatus

Service enquiry