On this page
The volume between two stroke marks
If only part of the stroke is used, give the start and finish marks. For several syringes of the same size, retain an individual identifier or batch sampling record so a result can be traced to the piece measured.
A syringe that dispenses between two intermediate marks may need a different check from one that makes a complete stroke. For a driven syringe, identify whether the required result is the volume from a specified travel or the volume delivered over a timed interval. The latter also involves the drive and its timebase.
Laboratory syringe volume delivery
State whether the method requires contained volume, delivered volume or repeated delivery at a selected setting. Keep the matching plunger and any adapter with the syringe identity. This is a laboratory measurement enquiry; an instrument result should not be presented as approval of a syringe or procedure for patient administration.
Specifications and service checks
A laboratory syringe may be used for a complete stroke or for a smaller movement between two marks. Identify those start and finish marks and the fitted outlet. Any remaining volume in a removable needle or connection matters when comparing the delivered amount with piston travel.
- Range and configuration
- Hamilton’s 1000 Series Gastight table covers nominal sizes from 1 to 100 mL. Include the barrel capacity and end fitting. A single nominal size can be supplied with different fittings.
- Manufacturers and models
- Hamilton’s models 1001 and 1010 are examples of 1 mL and 10 mL laboratory syringes.
- What we check
- For a laboratory delivery syringe, compare the volume displaced at the marked positions. Identify the attached needle or outlet where it affects delivered volume.
- Reference equipment and accuracy
- Use an appropriate traceable volume or gravimetric reference. Very small deliveries need a balance and evaporation control that resolve the permitted error.
- When to repeat the check
- Review after replacing a barrel or plunger, or if travel becomes uneven or the outlet leaks. Identify the assembly that was actually tested.
Technical references Hamilton Gastight syringe specifications · NIST volumetric calibration procedures

