On this page
Conductivity and the TDS conversion setting
If the same device also reports conductivity, ask for the two modes to be identified separately. A change in the TDS factor can move the displayed result while leaving the conductivity reading unchanged. Record any temperature compensation setting alongside the comparison.
A meter set to a factor of 0.5 and one set to 0.7 can show different estimated TDS values from the same conductivity reading. Before sending both instruments for adjustment, check their configured factors and whether the displayed conductivity agrees under the same conditions.
Specifications and service checks
A TDS display often calculates its value from conductivity using a factor selected in the meter. Changing that factor can change the displayed TDS while the conductivity reading stays steady. Copy the factor and units from the instrument used for the original observation.
- Range and configuration
- Hanna gives the DiST 1 HI98301 a 0–2,000 ppm display range and a TDS conversion factor of 0.5. Record the unit and configured factor on meters that allow it to be changed.
- Manufacturers and models
- METTLER TOLEDO Thornton M300 and Hanna DiST testers illustrate different calculations based on conductivity.
- What we check
- Compare the applicable input or the full sensor chain with a traceable reference. For a TDS display, record the factor used to convert conductivity to its indicated TDS value.
- Reference equipment and accuracy
- Use suitable conductivity standards or a traceable electrical simulation for the electronic channel. For high-resistivity water measurements, contamination can dominate a liquid comparison.
- When to repeat the check
- Recheck after cell changes and after changes to the temperature compensation or TDS conversion factor. Trend results near the normal water-quality limit.
Technical references METTLER TOLEDO Thornton M300 specifications · Hanna DiST tester documentation