Temperature chamber qualification

A temperature chamber used for product testing needs an assessment over its usable workspace. What happens at the controller sensor and what happens around the test item are separate observations. The job can cover fixed temperatures, programmed changes or both.

Send an enquiry
Temperature chamber — equipment illustration
Equipment illustration
On this page

Conditions around the test item

Mark where the item sits and how much space it occupies. If its own heat output matters, label the operating state during the survey. That detail is essential when comparing a loaded test with an earlier empty-chamber map.

For example, if a test calls for a 30-minute hold, does that start when the controller first shows the setpoint or when the item’s sensor reaches the specified band? Those can be several minutes apart. If the item has its own sensor, identify its channel and whether the enquiry covers that sensor as well as the chamber.

Specifications and service checks

A specification calling for “one hour at temperature” needs a clear start event. State whether the time begins when the controller reaches its setpoint or when an independent probe at the load reaches the condition. The two starts can be several logging intervals apart.

Range and configuration
The ESPEC SU-242 is rated from −40 to +150 °C. Its model specification covers the chamber; the required uniformity and the volume to be surveyed come from the intended use.
Manufacturers and models
The SU and SH models in ESPEC’s benchtop series include different temperature and humidity configurations.
What we check
Map selected positions in the working volume at the agreed settings. Identify where the control sensor sits and report the difference between its indication and the reference readings.
Reference equipment and accuracy
Use a set of calibrated temperature sensors and a synchronised logger. The reference uncertainty should support the allowed variation within the test space.
When to repeat the check
Review after refrigeration or heater repair, a change of control probe or a significant change in loading. Compare repeat maps at the same locations.

Technical references ESPEC SU and SH benchtop chamber specifications · EURAMET climatic chamber calibration guide