Identify the sensor type, output scale, sensing location and which components stay together in service. If the plant reads a digital value as well as a current signal, specify which result needs to appear against the applied temperature.
For a linear transmitter configured to represent 0 to 200 °C by 4–20 mA, 100 °C corresponds to a nominal 12 mA. A simulated midpoint checks the electronics, while a temperature comparison with the assembled sensor also exercises its response.
Specifications and service checks
A motor-winding monitor may use one transmitter design for both Pt100 and thermocouple inputs. The selected sensor type determines the simulated input needed for the temperature check. A photograph of the channel setup is particularly useful after a spare unit has been fitted.
- Range and configuration
- List sensor type, lower and upper temperature values and output. A 0–100 °C transmitter and a 0–400 °C transmitter can both use the same 4–20 mA output.
- Manufacturers and models
- Emerson Rosemount 644 is an example of a transmitter that accepts resistance and thermocouple inputs and can report a 4–20 mA signal.
- What we check
- Compare simulated input and electrical output for the transmitter, or compare the complete probe-transmitter system at actual temperatures. Identify which result the plant needs.
- Reference equipment and accuracy
- Use traceable resistance or thermocouple simulation and an electrical reference for the output. A complete system comparison also needs a calibrated reference thermometer.
- When to repeat the check
- Review after changing sensor type, wiring compensation or output scaling. Retain the as-found configuration with the result.
Technical references Rosemount 644 Rail Mount Temperature Transmitter