Thermal mixer qualification

A thermal mixer can have a temperature set point and a separate motion or speed setting. Its removable blocks may also have different shapes. Tell us which block and locations support the temperature measurements recorded in your equipment file.

Send an enquiry
Thermal mixer — equipment illustration
Equipment illustration
On this page

Centre and corner temperatures in the block

For a mixer with a 24-position block, a reading near the centre and one in a corner can answer the question ‘does the whole block run at this temperature?’ A single block-sensor reading answers a narrower question. Mark the occupied positions, the block model and the speed used during the hold. On a timed programme, specify whether the clock starts at switch-on or after the block reaches its working temperature.

Thermal mixer block and mixing configuration

Measurements at selected positions distinguish the programmed setpoint from the condition within the loaded block. The report retains the configuration and loading arrangement so results can be compared when the block, vessel format or operating programme changes.

Specifications and service checks

A removable block can have its own identifying mark even though the temperature and speed displays belong to the base. If a second block is shared between units, record which block-base combination was fitted. This prevents later comparisons from combining observations made on different assemblies.

Range and configuration
List the fitted block, its identifier and the instrument’s own temperature-indication range. A block’s well count describes the hardware and is separate from a temperature specification.
Manufacturers and models
Eppendorf lists the ThermoMixer C base as catalogue number 5382 000.015 for its 220–240 V version. Record the SmartBlock fitted on the service date alongside the base identity.
What we check
For an equipment service record, identify the standalone temperature reading or detachable sensor and its associated block. The result should name the particular channel compared.
Reference equipment and accuracy
Use a traceable temperature reference with documented uncertainty for that channel. Keep the reference and any manufacturer service report linked to the fitted block.
When to repeat the check
Review the measurement record after a block exchange or input repair. Different blocks on the same base may have separate service histories.

Technical references Eppendorf SmartBlock catalogue · Thermo Fisher VeritiPro specifications · Eppendorf ThermoMixer C and SmartBlock identification

Enquire about Thermal mixer