On this page
Switching on heating and resetting on cooling
For an adjustable thermostat, list the settings used in service and the acceptable switching band. A record of the rising and falling transitions at each setting is more informative than a single reading taken while the contact is already closed.
A thermostat may open a contact at one temperature while heating and close it several degrees lower during cooling. Retaining both numbers shows the operating band. If your process specifies separate tolerances for the two transitions, compare each switching temperature with its own limit.
Specifications and service checks
If a heating enclosure keeps cycling after the dial is moved, record the actual switching temperatures on warming and cooling. The gap between them is the control differential. Identify whether the service request concerns the marked dial setting, the switching thresholds or both.
- Range and configuration
- Danfoss gives −5 to +35 °C for the KP 68 setting range and −30 to +15 °C for the KP 61. The contact differential and reset arrangement should be checked for the actual type number.
- Manufacturers and models
- WIKA’s TSD-30 is an electronic temperature switch. The Danfoss KP family includes thermostats with mechanical sensing elements.
- What we check
- Compare the sensor with an independent thermometer while observing the contact. Record the trip and reset temperatures and identify the contact tested.
- Reference equipment and accuracy
- Use a stable temperature source, calibrated reference probe and contact indicator. Probe position and response time affect the observed switching temperatures.
- When to repeat the check
- Check after a setting adjustment or replacement sensor. Select the interval against the allowable trip-point error and the importance of the controlled equipment.
Technical references WIKA TSD-30 operating instructions · Danfoss KP thermostat data