Transfer pipette calibration

Transfer pipette may denote a calibrated single-mark glass pipette or an ungraduated disposable dropper. The two do not define a delivery in the same way. Identify the construction, marked capacity and any engraved delivery convention before agreeing the job.

Send an enquiry
Transfer pipette — equipment illustration
Equipment illustration
On this page

Volume markings and how the vessel is used

For a marked volumetric pipette, give the stated delivery and reference temperature. For an unmarked dropper used for a defined transfer, send us your delivery method, tolerance and lot sampling requirement. We can quote the check against the transfer you actually make.

If a purchasing record calls an ungraduated disposable dropper a 1 mL transfer pipette, verify what the stated capacity means on the manufacturer's documentation. A maximum bulb capacity is a different quantity from a defined delivered volume assessed with a specified transfer method.

Specifications and service checks

Some transfer pipettes are marked for a stated volume while others simply move drops. If the printed catalogue calls it a transfer pipette, inspect the actual volume mark and delivery convention. For an unmarked dropper, identify the physical or functional feature your procedure needs checked.

Range and configuration
For a marked glass transfer pipette, a 25 mL bulb pipette is one example of a one-volume instrument. An ungraduated disposable dropper needs a separately agreed delivery requirement. Identify which type is being purchased or checked.
Manufacturers and models
BRAND BLAUBRAND includes bulb and graduated glass pipettes. Corning’s Stripette is an example of a serological pipette.
What we check
Measure the amount delivered under the drainage convention marked on the instrument. Record which marks were used and whether blow-out is specified for that design.
Reference equipment and accuracy
A traceably calibrated balance and thermometer support a volume comparison. Choose resolution and evaporation control for the smallest delivery under test.
When to repeat the check
Identify glass pieces individually where practical. For disposable serological or transfer pipettes, follow the lot-control requirements used by the laboratory.

Technical references BRAND volumetric glassware testing guide · Corning Stripette serological pipettes · NIST volumetric calibration procedures · Corning 4488 10 mL Stripette · BRAND one-volume bulb pipettes