On this page
Anvil shape and the tube wall
Record the internal anvil shape and the usual wall thickness, tube bore and distance from the end. When comparing results with an older instrument, check the anvil geometry as well as the readings. Different contact forms can locate on different points of a curved wall.
The anvil may fit inside a straight tube yet touch the bend of a formed elbow before it reaches the wall to be measured. Send a section drawing of the tighter opening and mark the contact point. This shows whether the frame and anvil have clearance at the working position.
Specifications and service checks
A tube wall read near its cut end can be thicker locally because of a burr. Mark the distance from the end and the angular position if the job concerns wall variation around a formed tube.
- Range and configuration
- Mitutoyo’s 115-302 measures 0–25 mm with a 0.01 mm graduation and a stated accuracy of ±3 µm. Its cylindrical anvil reaches inside tubing, so both wall thickness and minimum tube bore matter.
- Manufacturers and models
- Mitutoyo 115-302 is a mechanical tube micrometer with a cylindrical anvil. The family also includes other anvil shapes.
- What we check
- Compare the indication at several wall thickness equivalents with the anvil contacting the assigned reference surface as it does in use. Inspect the inner anvil for wear at the working line.
- Reference equipment and accuracy
- Use traceable length standards or suitable wall-thickness artefacts that permit correct seating without bridging. Include repeatability of line or point contact in the uncertainty.
- When to repeat the check
- Check again after a tube catches the slender anvil or the normal zero moves. A reference near the routine wall thickness is especially useful between calibrations.
Technical references Mitutoyo Tube micrometers