Ultra-low-temperature freezer qualification

An ultra-low freezer survey needs references suitable for its actual storage temperatures. The door display, independent alarm probe and temperatures inside the storage boxes can describe different locations. List them separately when specifying the job.

Send an enquiry
Ultra low temperature freezer — equipment illustration
Equipment illustration
On this page

Rack loading and recovery after access

Record the arrangement of inner doors and racks, the occupied compartments and the intended monitoring period. If a recovery check is required, agree how recovery will be judged and keep that result separate from the steady temperature distribution.

Two reports from the same freezer can show different extremes if one run included filling a rack and the other took place overnight with the door closed. A brief event log, marked with the same clock time as the loggers, lets your quality team find that change in the traces. Note any difference between the freezer panel clock and logger time before comparing the plots.

Specifications and service checks

If the installed sensor is fixed high inside the cabinet while the stores logger is packed in a bottom drawer, label each position and record loading conditions. Review both time traces around a known door opening before interpreting a difference as an offset between instruments.

Range and configuration
Eppendorf quotes a −86 to −50 °C setting range for its CryoCube F740hi family. Specify the freezer’s working temperature and the shelf or compartment that the comparison needs to represent.
Manufacturers and models
Eppendorf’s CryoCube F740hi and F740hiw are ultra-low temperature freezer models. Use the exact model suffix when identifying the refrigeration configuration and equipment manual.
What we check
Compare the unit’s indication with independent readings in the storage space. Record shelf position, loading and any defrost or door events during the agreed observation.
Reference equipment and accuracy
Use probes whose calibration extends to the lowest temperature required. Lead routing and settling time can affect readings, especially in an ultra-low unit.
When to repeat the check
Check again after a refrigeration repair, control probe replacement or persistent monitor alarm. Compare drift near the normal storage temperature.

Technical references Liebherr laboratory freezer manual · Eppendorf CryoCube F740hi manual