On this page
Drive checks and the rotor record
List speed, timer, temperature and any other monitored indications as distinct requested checks. A rotor history or inspection record should remain traceable to the rotor itself and separate from the calibration of the drive or its display.
For a unit whose last rotor change occurred between two service visits, identify the rotor named in each record before comparing calculated centrifugal force. Speed results and the rotor-specific calculation have different dependencies, even when the instrument ID and installed drive have stayed the same.
Specifications and service checks
A drive speed and rotor service history may be maintained under different record numbers. When asking for a speed comparison, give the installed rotor designation and the speed unit on the display. Keep rotor life or inspection records linked to the rotor’s own serial number.
- Range and configuration
- Beckman Coulter gives 100,000 rpm as the maximum for the Optima XPN-100. A specific rotor or accessory can impose a lower limit.
- Manufacturers and models
- Eppendorf MiniSpin plus and Beckman Coulter Optima XPN-100 illustrate widely different speed classes.
- What we check
- For instruments with indicated speed, request a comparison using the manufacturer-approved service method. A fixed-speed mini unit can be reported at its operating speed.
- Reference equipment and accuracy
- Speed traceability must suit the range and the drive’s authorised measurement arrangement. A regular handheld tachometer may not suit a sealed high-speed system.
- When to repeat the check
- Check the speed indication after drive or sensor service. Keep the rotor service-life record separate from the calibration date of the speed measurement.
Technical references Eppendorf MiniSpin documentation · Beckman Coulter Optima XPN-100