On this page
Distance and transit time in the velocity result
For a 0.30 m path and a measured time of 75 µs, the calculated velocity is 4,000 m/s. An error in the entered distance changes that figure even if the timer reads correctly. Send the transducers and cables used together, the reference bar and the zero correction in use. For a site comparison, tell us whether the probes are on opposite, adjacent or the same face.
Ultrasonic pulse velocity tester
Identify the transmitting and receiving probes, reference bar and cable arrangement. State which time indication or reference comparison is required. Measurements made on concrete also depend on the selected path and conditions, so an equipment check should remain separate from a judgement about concrete quality or strength.
Specifications and service checks
A direct path through opposite faces and an indirect path along one surface do not share the same probe geometry. Label the actual contact positions on a sketch when comparing transit times measured on a concrete member.
- Range and configuration
- The Pundit Lab package includes two 54 kHz transducers. Velocity depends on path length and transit time, each with its own uncertainty.
- Manufacturers and models
- Screening Eagle’s Pundit Lab and Pundit Lab+ use a separate transmitter and receiver.
- What we check
- Verify transit time with the manufacturer’s characterised reference and the paired transducers. If velocity is reported, independently establish the length used in its calculation.
- Reference equipment and accuracy
- Use the identified reference rod with its assigned value, a traceable time reference as appropriate and a length standard for the path. Document the supplied cables.
- When to repeat the check
- Repeat the verification after transducer or cable replacement and before a field series when required by the inspection method.
Technical references Screening Eagle 7. Part Numbers and Accessories