Optical reference values and cell positions
When a reader has a multi-position cell changer, a shift that follows one position is worth noting separately from one seen in every cell. Tell us which positions and wavelengths need checking. A reference filter or material has an assigned response under stated conditions. The filter identity and measured absorbance belong together, with separate results for wavelength, stray light or resolution when you request them.
UV visible spectrophotometer performance
Identify the reference filters or solutions and the performance checks in its protocol. Wavelength accuracy, photometric response and other optical characteristics are separate records. A result obtained for one mode or cell arrangement should not silently establish every configured measurement mode.
Specifications and service checks
After blanking, a filter that transmits half the reference light at the selected wavelength gives 50% transmittance, or about 0.301 absorbance. The conversion is A = −log₁₀(T), with T entered as 0.50. If your worksheet accepts readings in %T, specify the %T limits and working wavelength on the order so the laboratory compares the result on the scale you use.
| Question | Quantity compared |
|---|---|
| Does a spectral feature appear in the right place? | Observed wavelength against an assigned reference band, in nanometres, at the selected bandwidth. |
| Does the detector report the correct light transmission? | Indicated transmittance or absorbance against a reference assigned for that quantity and wavelength. |
- Range and configuration
- Agilent specifies 190–1,100 nm for the Cary 60. The instrument’s usable absorbance also depends on wavelength, optical bandwidth and the installed accessory.
- Manufacturers and models
- Agilent’s Cary 60 can be fitted with different sample holders and accessories, which should be identified on the service request.
- What we check
- Check wavelength indication and photometric response separately. Use assigned spectral features for one and suitable certified absorbance or transmittance references for the other.
- Reference equipment and accuracy
- Select certified references with values at the actual wavelengths and bandwidths used. The certificate should identify the reference, optical path and uncertainty.
- When to repeat the check
- Review after lamp, detector or optical path service. Interim checks near the working wavelength can show a shift before a scheduled annual review.
Technical references Agilent Cary 60 specifications · NIST SRM 2034 reference certificate · NIST wavelength and photometric references · IUPAC definition of absorbance