Tester output and interrupter test settings
Give the exact tester model, test mode and quantities from the maker’s procedure that require calibration. If only the tester is supplied, keep its results separate from a switchgear or interrupter test performed later at the installation.
Vacuum bottle testers
Record the instrument's output type, voltage indication, timing and leakage or pass/fail functions. The word vacuum here does not mean that it measures chamber pressure. Send the manufacturer procedure and required checks for assessment before arranging the job; a vacuum-gauge calibration request would address a different measurement altogether.
Specifications and service checks
A switchgear records system may list the breaker pole under test, while the tester stores only the programmed sequence. Connect the two identifiers in the enquiry, along with the function in question, so a suspected timing or voltage problem is separated from the switchgear result.
- Range and configuration
- For this application the relevant settings are test voltage and application time, together with the defined outcome of the test.
- Manufacturers and models
- Megger VIDAR assesses vacuum interrupter integrity using a specified DC test voltage. It does not display chamber vacuum in pressure units.
- What we check
- Verify the tester's output and indication against suitable references. Keep the tester's verification separate from the result for an individual interrupter.
- Reference equipment and accuracy
- Use a rated calibrated high-voltage reference and traceable timer, plus any manufacturer-defined load or functional verification accessory.
- When to repeat the check
- Review after output repairs or lead replacement and follow the interrupter manufacturer's limits for the asset test.
Technical references VIDAR - Vacuum interrupter tester