Vacuum range endpoints and signal output
List the lower and upper range values, pressure reference and output signal. Where two vacuum ranges share one controller, identify which transmitter drives each channel. Include any pressure at which the system switches from one range to the next.
If a vented chamber reads zero on the controller, check which scale is displayed. The panel may be showing vacuum relative to atmosphere or a percentage. Record the transmitter’s actual output beside the configured reference before comparing the panel with an absolute-pressure gauge.
Specifications and service checks
A vacuum transmitter that feeds a recorder may have a logarithmic pressure output. A halfway electrical signal need not represent halfway between its pressure endpoints. Providing the maker’s conversion curve helps distinguish an incorrect recorder scale from a sensor error.
- Range and configuration
- Identify the pressure decades and the signal conversion used by the controller. A linear signal-to-pressure assumption can give the wrong result on a logarithmic output.
- Manufacturers and models
- INFICON PCG550 is a combined Pirani/capacitance gauge with a selectable electrical output.
- What we check
- Compare the response with a vacuum standard over the operating region for the specified gas. Include both raw output and interpreted pressure if the loop is in question.
- Reference equipment and accuracy
- Use vacuum transfer standards covering the requested pressures and a calibrated electrical readout for an analogue signal.
- When to repeat the check
- Check again if filament, sensor head, gas service or configured output characteristic changes.
Technical references PCG550, PCG552, PCG554