The settings used for routine transfers
Copy volume settings with their units exactly as read on the instrument. If the pipette has a locking dial, note any setting or locking problem seen by the user. Keep results obtained before service separate from readings taken after a repair or adjustment.
A buyer who uses a 20–200 µL pipette only at 25 µL and 150 µL can flag those routine settings in the enquiry. The laboratory can then distinguish process-specific points from any other points required by the service procedure or acceptance specification.
Specifications and service checks
If 20 µL and 80 µL are the two routine settings on a 100 µL pipette, request both explicitly alongside any other selected points. Each represents a separate operating decision, and the effect of a given delivery difference is proportionately larger at the lower setting.
- Range and configuration
- The Gilson P200 M example spans 20–200 µL. Record both volume limits and the tip specified for the particular pipette. One increment on the volume display is not its allowable delivery error.
- Manufacturers and models
- Gilson PIPETMAN Classic and PIPETMAN M are examples of mechanical and electronic families. The complete model code identifies the volume range.
- What we check
- Select settings representative of the lower, middle and upper parts of the intended working range under the applicable method. A nominal-volume result alone leaves the low end uncharacterised.
- Reference equipment and accuracy
- Choose a calibrated balance with resolution and uncertainty suited to the smallest delivery. Account for water temperature, evaporation and air buoyancy when converting mass to volume.
- When to repeat the check
- Use the frequency of use, local checks and previous results to set recall. Give a leaking or repaired pipette a fresh check before returning it to the bench.
Technical references Gilson PIPETMAN Classic guide · Gilson PIPETMAN M specifications · ISO 8655-6 gravimetric volume determination