On this page
Contact face and vernier graduation
Identify the contact face, the smallest vernier graduation and any clamp position used for readings. If zero is established on a setting block above the plate, give the block value. The result should make clear whether displayed readings are heights from the plate or deviations from that setting.
If a stop block establishes a reference height above the surface plate, the operator may record heights as differences from that block. Give the block height and note this convention so the certificate and production record can be compared without an unexplained offset.
Specifications and service checks
For a scribed line referenced 40 mm above a machined base, placing the part on parallels changes its height above the plate. Show the stop, parallel and part on the same sketch if the gauge is set from the plate scale.
- Range and configuration
- Mitutoyo’s 514-102 covers 0–300 mm with 0.02 mm graduations and a stated accuracy of ±0.04 mm. Read the specified accuracy separately from the size of one vernier division.
- Manufacturers and models
- Mitutoyo 514-102 is an example from its Series 514 adjustable-main-scale vernier height gauges.
- What we check
- Place the gauge on a verified surface plate. Compare its vernier reading against reference heights, noting any shift when the carriage is clamped at a position.
- Reference equipment and accuracy
- Use assigned steps or calibrated gauge blocks and check the plate beneath both the reference and the base. Select the reference uncertainty against the height tolerance on the job.
- When to repeat the check
- Inspect the base and scriber for burrs. Recheck the zero and a working height if the adjustable scale or scriber has been disturbed.
Technical references Vernier Height Gage with Adjustable Main Scale - Mitutoyo · Mitutoyo Height Master