On this page
Indentation force and the optical reading
Identify each force used, the objective or camera and the evaluation mode. If you compare a manual reading with an automatic one, request results for those modes explicitly. Reference block values should include their test condition, especially when the same machine covers several force ranges.
Two shifts may use the same test force but read the indent with different objectives. If their hardness results disagree, keep both diagonal measurements and the objective used on each shift. That gives the laboratory an optical reading to compare as well as a calculated hardness value.
Specifications and service checks
A Vickers result uses both diagonals of the impression. If one corner is poorly defined, keep the two individual readings and the objective used. The final hardness number alone leaves no way to see why operators selected different edges.
- Range and configuration
- Buehler lists Wilson VH1102 and VH1202 microindentation models for loads from 0.01 to 2 kgf. Record the exact test force and the objective used to measure the diagonals.
- Manufacturers and models
- Wilson VH3100 is a Vickers and Knoop tester. The chosen indenter and reporting scale identify the method.
- What we check
- Make and read impressions under the specified force and dwell, using a certified block for that HV designation. Check both diagonal readings and the optical scale separately when their performance is in question.
- Reference equipment and accuracy
- Choose a Vickers block certified at the stated test force and a calibrated optical scale. At low loads, focus, diagonal resolution and surface preparation become especially important.
- When to repeat the check
- Check the optical scale and a reference block after changing an objective, indenter or camera settings. Track results at the force routinely used.
Technical references Wilson VH3100 Vickers and Knoop Hardness Tester - Buehler · Buehler Micro Vickers hardness tester reference