Matching logger traces to the room plan
For a chamber with several programmes, state which conditions need mapping. Retaining the position codes on both the drawing and the time traces gives the next engineer a way to repeat a survey after the room layout changes.
If two loggers are marked ‘back wall’, the labels may be indistinguishable once the instruments have been removed. A position such as ‘rack B, upper shelf, door end’ can be found again on a plan and in the data file. For an interconnecting room, give the anteroom and the main working space separate position codes.
Specifications and service checks
A new pallet position near a door can extend the occupied storage zone beyond the earlier survey. Add its height and floor location to the plan, together with the existing monitoring sensor. This shows which readings come from newly occupied space and which can be compared year to year.
- Range and configuration
- For one Weiss Technik walk-in chamber package the maker lists climate testing from 10 to 95 %RH at +10 to +90 °C. The selected enclosure and climate envelope determine which paired conditions are available.
- Manufacturers and models
- Weiss Technik supplies configurable walk-in chambers with optional climate equipment.
- What we check
- Agree on a position plan for the working volume, document occupancy and compare the fixed controller with distributed independent measurements.
- Reference equipment and accuracy
- Use calibrated loggers with synchronised clocks. Select probes for the requested temperature and humidity and include their position on the measurement record.
- When to repeat the check
- Review a previous map after changes to shelving, air distribution or doors. An annual interval is meaningful only alongside a stable configuration and satisfactory checks.
Technical references Weiss Technik walk-in chambers · EURAMET climatic chamber calibration guide