Walk in freezer qualification

A walk-in freezer can have several evaporator areas, doorways and storage heights. A wall display at one end cannot represent all of them. A useful survey follows the storage layout rather than dividing the floor alone into equal squares.

Send an enquiry
Walk in freezer — equipment illustration
Equipment illustration
On this page

Shelf heights, access and defrost times

Share the normal access times, defrost schedule and rack elevations before the visit. A temperature excursion can then be read against real events, and a survey done after adding a rack can be compared with the earlier arrangement.

If the freezer has more than one evaporator or a scheduled defrost cycle, include the cycle times with the monitoring record. A rise observed near one evaporator can then be located in both space and time instead of being mistaken for a uniform rise throughout the room.

Specifications and service checks

A defrost event and a loading event can both appear as a warm peak on a logger. Record the defrost times and the actual door-opening times against the same clock. The traces from the door and rear racks can then be interpreted in the context of the event.

Range and configuration
Give the design storage band, room dimensions and positions of the fixed sensors. A logger used to map the room must have a calibration that reaches the lowest requested point.
Manufacturers and models
Viessmann offers TectoCell Standard Plus enclosures with panel thicknesses including 100, 120 and 150 mm. For a freezer, use the plant specification to identify its actual storage temperature and acceptable variation.
What we check
Compare fixed sensors with independent records at marked positions. Identify door activity, defrost periods and loading in the record so unusual readings can be traced.
Reference equipment and accuracy
Use calibrated loggers suitable for the lowest intended temperature. Reference uncertainty, response time and placement affect the interpretation of short excursions.
When to repeat the check
Review after changing evaporators, shelving or door arrangements. Use the previous mapping and continuous monitor records to set the next check.