Contact edges on the working slots
On a slotted plate, mark the slots in regular use and the contact edges that control entry. On an indicating device, include the anvils and normal contact orientation. Request results in length units if the gauge is used to check a drawing dimension.
Wire gauge openings and their marked sizes
Record the numbering system and marked sizes, and send the dimensional table used by your inspection procedure. The number on a slot and the opening to be measured should remain linked on the enquiry and report.
The permanent-facility section of CC-4643 lists wire-gauge linearity from 0.19 mm to 10 mm using a vision measuring machine. Select the openings or features needed within that entry and give the acceptance limits. For another gauge construction or a dimension outside that listed range, include the drawing and required features so the work schedule identifies the separate measurement requirement.
Photograph a chipped or worn opening and identify where the wire normally contacts the gauge. Avoid treating a diameter measured on a sample wire as the calibration result for the gauge itself. For several plates or discs, use a separate asset number for each; similar markings can otherwise make the returned results difficult to reconcile.
Specifications and service checks
A plate marked with SWG and another marked AWG can associate different diameters with the same gauge number. Record the series on the holder and the marked hole or slot before comparing the measured wire against a job card.
- Range and configuration
- Starrett’s 281 covers American wire gauge sizes 0 through 36. The equivalent diameter for a gauge number depends on the named wire standard.
- Manufacturers and models
- Starrett 281 is an American Standard wire gauge. State the gauge system if a different plate is used.
- What we check
- Measure the slot or opening width at the working contact and compare it with the corresponding size table. Record wear or damage at the mouth separately.
- Reference equipment and accuracy
- Use a calibrated vision system or suitable dimensional comparator with uncertainty small enough for the narrowest slot. A wire number alone is not a traceable length.
- When to repeat the check
- Inspect the narrow openings frequently if abrasive wire is drawn through them. Recheck after burr removal.
Technical references Starrett 281 American Standard Wire Gage, 0-36 Range · CC-4643 calibration scope, page 21