+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Yarn strength tester testing

A yarn strength tester measures force as yarn is extended between grips. Breaking force, extension and calculated tenacity are different reported results. Gauge length and the loading or extension setting help define what the machine measured.

Send an enquiry
Yarn strength tester — equipment illustration
Equipment illustration
On this page

Breaking force, gauge length and tenacity

Two yarns can break at the same force and still have different tenacities because their linear densities differ. If your report shows cN/tex, identify the tex value used in the calculation as well as the load-cell reading. For the machine verification, give the force range, initial grip separation, test speed and grip type. These let us tie the strength result back to the force and movement checks.

Yarn strength tester force and extension

Specify whether force, extension, speed or several quantities need records. Gripping and specimen preparation remain part of the laboratory’s test process. RTS checks the force, extension and other specified instrument measurements separately from the yarn-test method.

Specifications and service checks

If a yarn slips through a grip, the recorded extension includes movement at the grip. Include the gauge length, grip style and where the break or movement occurred when deciding whether to check force, travel or the grips.

Range and configuration
The useful span depends on the fitted load cell. Also record jaw separation and crosshead travel speed, especially when results include elongation.
Manufacturers and models
James Heal’s Titan family and SDL Atlas tensile equipment can be configured for yarn work. The SDL Atlas catalogue identifies the 5ST single-end yarn strength tester.
What we check
Verify indicated tension over the working portion of the load cell. Check travel and speed independently. Examine the grips if successive tests show slipping.
Reference equipment and accuracy
Use a traceable force transfer standard for the direction of loading, plus length and time references for motion. Match force uncertainty to the lowest point requested.
When to repeat the check
Review after overload, a load-cell replacement or a grip rebuild. Keep the load-cell serial number with the verification record.

Technical references James Heal product reference · SDL Atlas product reference · SDL Atlas yarn testing catalogue

Enquire about Yarn strength tester

Service enquiry