On this page
Match each rod to its working depth
Record the head serial number, base width, display resolution and the range marked on each rod. If the rods have individual marks, retain them in the enquiry. Where rods carry only their nominal ranges, describe how stores distinguishes the pieces. For a single-rod instrument, give its usable range directly.
Use the depths called up on the drawing
A counterbore inspected near its nominal depth uses only a small part of the scale. Give that dimension and the tolerance required by the drawing. For a stepped part, list the depths separately, especially when one crosses into another rod range. State if the operator sets zero against a particular artefact before inspection.
A rocking base changes the measurement
A base that bridges a narrow recess can sit differently when it is turned through 90 degrees. If your depth readings change with orientation, sketch the shoulder and the area the base rests on. Note which rod was installed and whether you saw the same change with another rod. We can then examine the base, the rod tip and the head as parts of the same tool.
Depth micrometer
On a depth micrometer with interchangeable rods, give each rod's nominal range. A correct zero with the shortest rod says little about where another rod seats when fitted to the same head.
Two interchangeable rods can cover neighbouring ranges but have different zero offsets. For a 32 mm recess, record the rod installed and the datum used to set it. A check on the shorter rod alone will not explain that reading.
- Range and configuration
- The Mitutoyo 129-131 is a 0–4 in interchangeable-rod model graduated in 0.001 in. The short travel of the screw head is only part of the set’s total measuring range.
- Manufacturers and models
- Mitutoyo’s Series 129 mechanical and Series 329 digital depth micrometers use interchangeable rods. The full order number identifies the set.
- What we check
- Check the base and zero before comparing each rod at identified depths. Include intermediate screw positions so a correct zero on every rod does not hide travel error in the head.
- Reference equipment and accuracy
- Use calibrated depth steps, for example a depth micro checker, and gauge blocks covering intermediate points. A warped base can dominate a comparison even with high-grade blocks.
- When to repeat the check
- Keep the rods associated with their head. After changing a rod seat or dressing the base, compare the affected rod combinations again.
Technical references Mitutoyo Depth micrometer reference · Depth Micro Checker - Mitutoyo