Output load, amplitude and polarity
List the functions actually programmed on the bench. For AC include frequency, amplitude convention and load. For DC specify polarity, level and load current. Identify any value read back by the source separately from the value measured at its terminals.
If a bench applies a positive DC bias and later switches to a sinusoidal test signal, list the two programmed conditions on separate lines. Include the offset if the AC signal itself carries one. That distinguishes two operating modes from a waveform containing both components.
Specifications and service checks
A design bench may switch between positive and negative DC and a zero-offset AC signal. Note whether the load stays connected during the mode change and which settled outputs matter. A bipolar DC requirement can be missed if an order lists only the largest positive value.
- Range and configuration
- Many variants in this series have 0–150 V and 0–300 V ranges. The current available changes with range, output power and the load.
- Manufacturers and models
- Chroma's 61500 series includes programmable AC sources with selectable output modes. Confirm which modes are fitted on the specific unit.
- What we check
- Compare steady DC of each required polarity and AC RMS output at stated frequencies. In mixed AC/DC operation, identify the offset and alternating component separately.
- Reference equipment and accuracy
- Use a traceable true-RMS voltmeter or power analyser with adequate bandwidth and a reference frequency counter. Choose a rated test load for load-regulation work.
- When to repeat the check
- After an output-stage change, check delivered settings again. Repeated results at the usual load are useful when reviewing the interval.
Technical references Chroma High Performance Programmable AC Power Source