Test volume and antenna separation
Identify the chamber layout, intended test method and measurement volume. A validation result for one antenna separation or frequency band should not be applied to another by assumption. Include any absorber, turntable or antenna changes since the last characterisation.
Anechoic chambers and the measurement system
Chamber performance, antenna behaviour and the instruments used inside it are different parts of the system. Provide the existing site-validation report and describe changes to absorbers, positioners or cabling. RTS assesses the chamber under the stated geometry and frequency conditions, with component calibration and chamber performance recorded separately.
Specifications and service checks
Moving a large turntable or replacing absorber can alter a chamber setup while its dimensions stay unchanged. Identify the usable test volume and the configuration in service when the change was made, so a site performance check can address that arrangement.
- Range and configuration
- The usable volume, distance and frequency range describe the site. A point field-strength reading alone cannot characterise the entire test space.
- Manufacturers and models
- ETS-Lindgren FACT chambers illustrate EMC test spaces specified for a measurement distance and frequency band. An antenna-pattern chamber has different acceptance measurements.
- What we check
- Verify the site characteristic for the intended work, such as attenuation, site VSWR or field uniformity, at the specified positions and polarisations.
- Reference equipment and accuracy
- Use calibrated antennas, receiver or network analyser, field probes and traceable positioning appropriate to the site test. State the setup and uncertainty with the result.
- When to repeat the check
- Revalidate affected conditions after absorber work, turntable replacement or a change in room geometry.
Technical references ETS-Lindgren FACT 5 EMC chambers