On this page
Polarisation, mounting and antenna connections
Give the model, frequency range, polarisation and intended mounting. Identify any balun or permanent cable belonging to the antenna. If your field-strength calculation uses cable loss separately, define the reference connector so its loss is accounted for once.
Antennas used as receiving references
State whether you need antenna factor, gain, impedance or another result, with orientation and geometry defined. Keep the original calibration data with the antenna, including any matching unit. This is a specialist reference enquiry; a receiver calibration does not assign a new antenna factor or establish the performance of its complete installation.
Specifications and service checks
For a survey antenna, an antenna-factor entry belongs to a frequency and its stated conditions. If records refer to two extensions or different antenna lengths, identify which arrangement was in place. The same serial can cover more than one setup.
- Range and configuration
- An EMC antenna may need antenna factor in dB/m; a communications antenna may need gain in dBi. Identify the quantity and frequency points required.
- Manufacturers and models
- ETS-Lindgren Model 3115 is a double-ridged guide antenna specified from 750 MHz to 18 GHz. Include the antenna's serial number and polarisation orientation.
- What we check
- Compare response with a calibrated reference antenna using a characterised site or suitable antenna calibration method. Specify orientation and the cable reference plane.
- Reference equipment and accuracy
- Use a calibrated reference antenna and receiver or network analyser. Site geometry, cable loss and mismatch all affect uncertainty.
- When to repeat the check
- Repeat after connector repair, bent elements or damage in transit.
Technical references 3115 Double-Ridged Guide Antenna - ETS‑Lindgren