+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Antenna analyzer calibration

An antenna analyzer can display impedance, return loss or standing wave ratio at its test port. A result at the end of a feed line can differ from a result at the antenna socket because the cable is then part of the measurement.

Send an enquiry
Antenna analyzer — equipment illustration
Equipment illustration
On this page

The impedance at the analyser port

Identify the connector, reference impedance and frequency points used on site. If open, short and load standards or a port extension are normally applied, describe that arrangement. Specify which displayed quantity is used for acceptance.

If an installer accepts an antenna at a stated return loss while another records VSWR, send the limit with its unit or ratio. A reading in dB and a VSWR figure should not be compared as though they were on the same scale. Both also need a defined test frequency.

Antenna analyzers

Include the calibration kit and adapters used to establish the working connection. A cable-and-antenna measurement involves a defined reference plane, so show where that plane is expected to be. RTS checks the specified analyzer functions with the port, options and reference plane identified.

Specifications and service checks

A measurement at the end of a long feeder includes the feeder and connectors. A measurement made at the antenna feedpoint answers a different question. Identify which position an installation team uses before specifying a return-loss comparison.

Range and configuration
Return loss is in dB, VSWR is a ratio and distance-to-fault depends on the cable's propagation velocity. Specify which functions will be reported.
Manufacturers and models
Anritsu Site Master S331E is a handheld cable and antenna analyser. Configuration and test lead define its working reference plane.
What we check
Check reflection measurements using known loads and mismatches. For distance measurements, use a line of known electrical length and record the velocity setting.
Reference equipment and accuracy
Use characterised RF standards and an appropriate verification kit. Keep the standards used to verify the result independent of the initial port adjustment.
When to repeat the check
Track connector wear and verify after test cable damage or replacement.

Technical references Site Master Compact Handheld Cable and Antenna Analyzer S331E

Enquire about Antenna analyzer

Service enquiry