Calibration slide calibration

A calibration slide carries a known pattern or ruled interval for checking an optical measuring system. The lines can be read through different objectives or over different parts of a field. Knowing only the overall glass size does not identify the usable reference.

Send an enquiry
Calibration slide — equipment illustration
Equipment illustration
On this page

Ruled lines, grids and circles on the slide

Provide the markings on the slide and the pattern used, such as a ruled scale, grid or circle. Give the interval or feature sizes required and their orientation. A result for line spacing on the slide can then be kept apart from a check on the instrument magnification.

Calibration slide markings

State which lines, grid features or other dimensions need assigned values. Include the previous certificate and the illumination arrangement where it matters to the method. A slide is a reference artefact; its assigned feature dimensions and an optical instrument's image scale need separate records.

Specifications and service checks

If 50 intervals span 0.50 mm, the mean pitch over that stretch is 0.010 mm. That average gives no individual spacing for the two lines used to sort a small feature. Mark those lines when local pitch matters.

Range and configuration
Graticules Optics S11 is an imperial stage scale spanning 0.005 inch in 0.0001-inch divisions. The object-plane value of an eyepiece division depends on the fitted objective.
Manufacturers and models
Graticules Optics supplies S-series stage micrometers. Give the engraved scale code alongside the microscope objective.
What we check
Compare the slide’s line centres on a calibrated optical measuring system. For an eyepiece, compare its divisions with a calibrated stage scale at each objective used.
Reference equipment and accuracy
Use a calibrated line scale or stage micrometer with uncertainty small relative to the line spacing. Retain the scale factor for each optical configuration.
When to repeat the check
Recheck eyepiece factors after objective, camera or magnification changes. For a slide, inspect line damage and review its separate recall history.