Capacitance meter calibration

Capacitance readings depend on the applied test frequency and, on some meters, on the series or parallel equivalent circuit selected. Lead and fixture capacitance can be significant at the low end of the range.

Send an enquiry
Capacitance meter — equipment illustration
Equipment illustration
On this page

Capacitance at the test frequency

Give the working capacitance values, frequency, signal level and normal fixture. If open or short compensation is used, record that state. For an instrument that reports loss alongside capacitance, specify both functions when the bench accepts components on both values.

A small capacitor measured directly at the terminals and then through a long lead can give two different results. Note whether fixture compensation was applied after the lead was connected. A change in the lead arrangement can reintroduce an offset, especially near the lowest range.

Specifications and service checks

Polarised capacitors used on a DC bus may have a specified test frequency for their nominal capacitance. Identify that frequency and the measurement fixture when comparing the meter with a component value. The nameplate value alone gives an incomplete comparison.

Range and configuration
A nominal 1 µF capacitor can give different results at 100 Hz and 1 kHz. For inductors, both frequency and test current can influence the reported inductance.
Manufacturers and models
Keysight E4980A measures impedance from 20 Hz to 2 MHz. The relevant settings include frequency, test level, fixture and series or parallel equivalent circuit.
What we check
Compare known capacitances at their stated test frequencies and levels, keeping the terminal configuration consistent. For small values, correct the open fixture before interpreting the reading.
Reference equipment and accuracy
Use calibrated impedance standards characterised at the chosen frequency, with appropriate open and short corrections for the fixture. Include loss or quality factor where specified.
When to repeat the check
Verify with a stable check component after changing the fixture or leads and investigate shifts before changing the adjustment.

Technical references Benchtop LCR Meters: E4980A

Enquire about Capacitance meter