Clean bench qualification

A clean bench can carry separate identifiers for the work enclosure, fan assembly and local indicator. The report should identify the bench as a complete installed unit and show where the readings refer to on the work surface.

Send an enquiry
Clean bench — equipment illustration
Equipment illustration
On this page

Marking positions on the working surface

Mark the usable area on a simple sketch and retain the orientation of each measurement position. A later reviewer can then compare the same position after a filter or fan change, even if the service company labels its grid differently.

The words ‘left’ and ‘right’ on a report can be reversed when someone reads a bench drawing from the opposite side. Mark the operator side on the layout and use it as the reference for position labels. This also helps your facilities team recognise a bench that moved rooms between reports.

Clean benches and laminar air flow cabinets

Record horizontal or vertical flow, filter layout, working area and the conditions required for the assessment. Do not substitute a biosafety-cabinet protocol solely because both enclosures have filters. RTS assesses the installed unit against the manufacturer’s requirements and the agreed test plan.

Specifications and service checks

When a clean bench is moved to a new room, an old installation record still describes the previous position. Mark the new location and any change to the fitted display or probe on the job request. This keeps instrument comparisons tied to the installation in use.

Range and configuration
Labconco lists nominal widths of 3, 4, 5, 6 and 8 ft for its Purifier horizontal clean benches. Width identifies the enclosure. For a calibration record, identify the installed indicator by its own serial number.
Manufacturers and models
Examples include the Thermo Scientific Heraguard ECO clean bench and Labconco Purifier horizontal clean benches.
What we check
Request verification against the clean-bench specification on the nameplate. Keep its installation record and the individual monitor’s calibration certificate together.
Reference equipment and accuracy
Ask for the test provider’s reference-instrument certificates and the equipment identity on the report. Use the maker’s specification for this class of bench.
When to repeat the check
Review service dates, installation moves and filter or monitor changes. A service visit and a completed performance verification should have separate records.

Technical references Thermo Scientific cabinet selection guide · Labconco Purifier clean benches

Enquire about Clean bench

Laminar air flow cabinet

The air direction can be horizontal or vertical. Record the orientation of the cabinet in the asset register so a sketch from its last inspection is read the right way round.

A service file may record a cabinet velocity reading as well as the serial number of the portable instrument used to survey it. Keep those two identifiers in separate fields. If the cabinet has been relocated, update the site and room before the next installation check.

Range and configuration
For a clean bench, identify the working width, airflow orientation and installed indicator. Dimensions alone do not state the cleanliness or protection classification.
Manufacturers and models
Labconco’s Purifier manual lists configurations including model 3889700. Quote the complete cabinet number and nameplate description, along with the identity of any separately serviced readout.

Shared with Clean bench — What we check · Reference equipment and accuracy · When to repeat the check · Technical references