On this page
Probe identity and measuring mode
For two thermometers that disagree, we first need the model and measuring mode of each. On a unit with a removable contact probe, identify the probe and display together. On an infrared model, the selected mode can change the indicated value. If the previous certificate names a test or reference mode, mention it when booking a repeat check so the readings can be interpreted on the same basis.
Clinical thermometers: identify the measurement mode
Include the required temperature points and any dedicated probe or reference accessory. Predictive or adjusted display modes need their own procedure and should not be treated as an ordinary industrial indication. RTS calibrates the thermometer in the manufacturer’s specified verification mode; product approval and clinical-performance studies remain distinct records.
Specifications and service checks
A unit used in predictive mode can have a separate test mode for comparison in a controlled source. Copy the model and identify the operating mode named in its service instructions. This avoids interpreting a prediction algorithm as the probe’s steady-state temperature reading.
- Range and configuration
- The Microlife MT 16C2 has a specified measuring interval of 32.00–42.99 °C. Check the intended function and model instructions before requesting a calibration.
- Manufacturers and models
- Microlife publishes the MT 16C2 digital thermometer and its measurement specifications.
- What we check
- Request an instrument comparison in the manufacturer’s specified test mode with the serial number and the scope of the check recorded.
- Reference equipment and accuracy
- Use a traceable temperature reference and comparator suited to the model’s resolution and mode. The user manual identifies its rated measurement conditions.
- When to repeat the check
- Follow the device maintenance instructions and the responsible institution’s equipment programme. Damage, cleaning damage or an anomalous check is a reason for reassessment.
Technical references Microlife MT 16C2 documentation