CMM probe calibration

On a coordinate measuring machine, changing from a short straight stylus to a long extension can alter the result obtained around a reference sphere. The same probe head may carry both tips, and the direction of approach matters. The head’s serial number alone does not identify the setup that inspected the part.

Send an enquiry
CMM probe — equipment illustration
Equipment illustration
On this page

Which stylus inspected the feature?

A ball tip on a straight stylus might reach an open face, while a star stylus is used to reach a bore from the side. A single entry labelled “CMM probe” loses that distinction. A probe list from your part programme normally identifies the tip diameters, extensions and orientations that matter. Include the reference sphere and the probe head. If it is the sphere itself that needs a dimensional result, mark it as a separate item.

The CMM probe and stylus assembly

Record the sphere or other reference artefact used in the machine's qualification routine. Changing a stylus length, orientation or module changes the configuration being checked. Keep routine probe qualification records available, but distinguish that software routine from independent assessment of the measuring machine and its probing performance.

Specifications and service checks

A new stylus ball may have the same diameter as the old one but a longer stem. Keep its qualification record with the actual build, including extensions and probe-head angle, before comparing old and new results on a deep side hole.

Range and configuration
A Renishaw TP20 can use five-way or six-way modules. Stylus length, tip size and the probing directions matter more here than machine axis travel.
Manufacturers and models
Renishaw TP20 uses interchangeable modules. Identify the module and stylus combination used for production.
What we check
Qualify each working stylus on the reference sphere, then check probing performance in the relevant directions. Keep the module, extension and orientation identifiable in the result.
Reference equipment and accuracy
Use a calibrated qualification sphere and a stable mount. The assigned sphere diameter and form uncertainty must support the probing performance being evaluated.
When to repeat the check
Requalify after a collision, a stylus change or a shift in the reference sphere. Trend the qualification residuals between machine verifications.

Technical references Renishaw TP20 introduction