On this page
Force ranges and the fitted platens
Give the indicated force ranges, the loads used to accept material and the platens installed. Where the machine stores a peak load at failure, identify that channel alongside the normal live indication. A force calibration should identify the loading direction and checked range.
A laboratory may test both small specimens and higher-load concrete cubes on the same frame. A check concentrated near maximum load may omit the lower band used for the small specimens, so put both working intervals on the request.
Specifications and service checks
One display can show both force and calculated stress. For a nominal 100 mm square face, 100 kN divided by 10,000 mm² gives 10 MPa. An incorrect specimen area changes the displayed stress even when the force channel reads correctly.
- Range and configuration
- The frame rating in kN gives the upper load. The useful lower end depends on the installed force channel and the lowest force for which verification is required.
- Manufacturers and models
- CONTROLS AUTOMAX PRO-M is an automatic compression machine family. Give the frame and readout identifiers.
- What we check
- Apply compression through the machine platens to a calibrated force-proving device. Compare force indications through the working portion of each active range.
- Reference equipment and accuracy
- Use a reference with a suitable compression classification and rated fittings. Verify alignment and document the uncertainty at low forces as well as at capacity.
- When to repeat the check
- Check again after hydraulic work, a load cell change or platen replacement. Daily zero checks help show when a more complete verification is due.
Technical references CONTROLS AUTOMAX PRO-M · ISO Force-proving instrument calibration