Polarity and load at the DC terminals
Identify the function, working levels, polarity and load connection. Where the instrument supplies a reference signal to an indicator, include both zero and the values used for the scale check. Specify remote sensing where it forms part of the normal connection.
A source set to 5 V may deliver that voltage at its terminals while a test fixture sees less at the end of its leads. Compare the same reference point at a stated load. A reading with the fixture disconnected can conceal the drop seen in the working circuit.
Specifications and service checks
A sensor simulator may use small negative and positive offsets about zero. Include the signed values used in the procedure. A set of points entirely on the positive side would miss a polarity or zero-crossing issue affecting half the simulator’s work.
- Range and configuration
- The E36312A has one 6 V/5 A output and two 25 V/1 A outputs. Full rated voltage and current may therefore differ across terminals on the same instrument.
- Manufacturers and models
- Keysight E36312A is a three-output DC example. Each output has its own rating and should appear separately on a calibration request.
- What we check
- Compare programmed and displayed voltage with actual output under agreed load. Current-limit or constant-current operation requires an independent current comparison.
- Reference equipment and accuracy
- Use a calibrated DC voltmeter, current reference or shunt, and a suitable load. A panel readback agreeing with its own setting does not establish delivered output.
- When to repeat the check
- Review after output repair or unexplained differences between the supply display and the bench's independent meter.
Technical references Expert Benchtop DC Power Supplies: E36312A