+91 70565 55582[email protected]Get a quoteRequest a call
RTS Calibration
Menu

Dial bore gauge calibration

A dial bore gauge is usually set to a known diameter before it goes into the bore. The inspector then looks for the point where the needle reverses as the head is rocked. A written reading of zero has meaning only if the setting diameter and the fitted contact are known.

Send an enquiry
Dial bore gauge — equipment illustration
Equipment illustration
On this page

Interpreting a departure from the setting ring

Suppose the gauge is set in a 50 mm ring and the inspector records a departure of 0.02 mm. The bore size can only be recovered from the record if the sign convention is known. Send the nominal diameter, fitted contact and displacement required on either side of zero. If two benches disagree, include how each records an undersize bore.

Transmission travel is different from bore size

A dial calibration tester checks how the bore gauge transmits movement to its indicator. The CC-4643 entry covers up to 2 mm of transmission travel at a least count of 0.001 mm in the permanent facility. Give the nominal bore diameter separately so we can identify the correct setting ring.

For example, a gauge set for a 50 mm bore may report small deviations around that setting. Write the 50 mm nominal setting on the instrument list, then state the separate movement interval over which readings are needed. Identify the head, indicator and installed contact combination. This gives RTS both the application size and the transmission requirement without confusing the two.

If the order also requires a result for the setting ring, list its identifier and requested dimensions separately. A certificate for that master and a transmission result for the gauge answer different parts of the inspection setup. Ask the quality contact to confirm which records the component inspection procedure needs before the purchase order is released.

Bore dial gauge

A bore dial gauge may have one dial head and several extensions in the same case. A result at the setting ring belongs to the head, extension and contact fitted at the time. Mark the combination used for each working diameter, especially if different machines share the set.

For a bore checked just below the mouth and again farther in, use the same ring setting and label each depth. A change between depths means something different from a shift seen at both depths after the gauge is reset.

Range and configuration
The anvil set determines the bore diameters the head can enter. The indicator then reads a much smaller change either side of the setting diameter.
What we check
Set the indicator on a ring matching the working bore and find the reversal point. Keep the dial with its paired gauge body during any further check of anvil motion or transmission.

Shared with Dial bore gauge — Manufacturers and models · Reference equipment and accuracy · When to repeat the check · Technical references

Enquire about Bore dial gauge

Specifications and service checks

When changing between 40 mm and 60 mm bores, the contact extension and spacing washers may change as well as the ring. Keep the washer combination with each target size so a return to the first job reproduces its travel.

Range and configuration
RTS lists bore-gauge transmission checks up to 2 mm of movement at 0.001 mm least count. That is indicator movement around a setting. The nominal bore diameter can be many times larger.
Manufacturers and models
Mitutoyo Series 511 comprises two-contact bore gauges. Record the fitted dial and anvil combination.
What we check
Set the gauge in a calibrated ring near the bore size and rock it through the reversal point. Check the indicator travel separately if the job uses more than a single setting.
Reference equipment and accuracy
Use rings with assigned internal diameters and uncertainties suitable for the bore tolerance. Identify the anvil, spacers and indicator used to obtain the result.
When to repeat the check
Repeat a setting check after rebuilding the anvil stack or changing the indicator. A tight spot while rocking the head is a reason to inspect the centring mechanism.

Technical references Bore Gages for Small Holes - Mitutoyo · CC-4643 calibration scope, page 16

Enquire about Dial bore gauge

Service enquiry

Service enquiry