Differential probe calibration

A differential probe measures a voltage difference between two inputs while both may sit at a common voltage relative to ground. Differential gain accuracy and rejection of a common signal are separate characteristics.

Send an enquiry
Differential probe — equipment illustration
Equipment illustration
On this page

Probe attenuation and input accessories

Include attenuation, input accessories, scope termination and the frequency range used. Identify whether the result concerns steady voltage, pulse amplitude or common-mode rejection as defined by the model. Keep the probe serial number associated with the channel used for the comparison.

A probe connected across two signals may see a small differential voltage even though both inputs are far from the instrument reference. An amplitude check with one input at ground does not establish rejection of their common signal. Name the quantity your comparison needs.

Specifications and service checks

A differential probe can reject a voltage common to both inputs imperfectly, with rejection varying by frequency. If a small differential signal appears contaminated by a switching waveform, describe the common-mode conditions as well as the desired signal amplitude.

Range and configuration
Identify attenuation, maximum working differential voltage and common-mode range. Common-mode rejection changes with frequency.
Manufacturers and models
Tektronix THDP0200 is a high-voltage differential probe. Its differential rating and common-mode rating are separate specifications.
What we check
Compare differential gain using known inputs. If rejection matters, apply a characterised common-mode signal and measure the residual response under defined conditions.
Reference equipment and accuracy
Use suitable calibrated differential and common-mode sources and a scope of adequate bandwidth. The fixture needs known symmetry for a rejection measurement.
When to repeat the check
Recheck after lead or tip replacement and review an unexpected change in common-mode pickup.

Technical references THDP0100/0200 and TMDP0200 User Manual