A digital level used on a machine bed reports inclination along one or two axes. A survey level used with a coded staff is a different instrument. On a machine level, note which base or V-groove is used and which direction the display treats as positive.

Send an enquiry
Digital level — equipment illustration
Equipment illustration
On this page

Reversal readings and displayed slope units

Supply the working slopes with the unit selected on screen, such as degrees or millimetres per metre. If the instrument is reversed during alignment, explain the reversal method. A reading taken relative to a stored zero needs that zero identified.

A level reading of 1 mm/m is a slope. On a 200 mm base that slope corresponds to a height difference of 0.2 mm between its ends. Keep the slope unit and base length together when using the reading to adjust a machine foot.

Specifications and service checks

If an electronic level displays a positive slope when its marked end is raised, turning it end for end reverses the slope sign. Note the end facing the machine datum when comparing alignment readings taken on different shifts.

Range and configuration
WYLER Clinotronic PLUS measures inclination up to ±60°. A high-sensitivity machine level may cover a much narrower span quoted in mm/m. Identify the display unit before choosing points.
Manufacturers and models
WYLER Clinotronic PLUS is a portable electronic inclinometer. A Leica Sprinter is instead a survey level used with a staff.
What we check
Measure several positive and negative inclinations on a tilt reference, including reversal near zero. For a differential system include the reference head and the normal connection between heads.
Reference equipment and accuracy
Use a calibrated angle table, level comparator or sine setup capable of reproducing the selected small slopes. Include the uncertainty of the mounting surface.
When to repeat the check
Review zero in both orientations after transport. Recheck after changing the electronic head, base or compensation settings.

Technical references WYLER Release 2.0 · WYLER Digital level reference

Enquire about Digital level