What starts and ends the timed interval?
Give the timing mode, trigger input, contact used and time interval from the process. Where a control output ends the cycle, identify that output event as the end point. Include units carefully on timers that change automatically between seconds and minutes.
For a sealing machine set to five seconds, the timing interval might run from a start contact to a timer relay, or all the way to jaw release. Identify the event your product specification uses before interpreting the stopwatch observation as a timer error.
Specifications and service checks
A controller might be set in decimal minutes. A setting of 0.10 min lasts six seconds, while 0:10 on a minutes-and-seconds display means ten seconds. Record the time format if a replacement timer appears to run the same programme at a different speed.
- Range and configuration
- A one-second display step matters on a short interval. Over one hour, a one-second gain or loss corresponds to approximately 278 parts per million.
- Manufacturers and models
- Omron H5CX programmable timers offer ranges from 0.001 s to 9,999 h across selectable settings. Identify the timer mode and installed contact or transistor output.
- What we check
- Compare the delay from the defined input event to the output transition. Include contact operating time if the contact, rather than the displayed countdown, controls the process.
- Reference equipment and accuracy
- Use a traceable frequency or time reference, appropriate counter and start/stop arrangement. Hand reaction time can dominate a short manually timed comparison.
- When to repeat the check
- Review oscillator drift against past results and recheck after timebase repair or any unexplained change in a routine timing check.
Technical references Stopwatch and Timer Calibrations (2009 edition) · Omron H5CX timer specifications · Seiko S141 stopwatch instructions