If your field sheet says ‘9 blows, 45 mm’, the depth scale supplied the 45 mm and the operator recorded the blows. Checking the rod graduations will address only part of that result. For an assembly check, send the cone and hammer details, the marked depth datum and the specified drop height. Identify any rods or extensions replaced since the last inspection.
Dynamic cone penetrometer components
Provide the component dimensions and mass requirements from the method, including any replaced parts. The apparatus checks concern these defined quantities. Interpreting penetration per blow as a soil or pavement property requires the separate field procedure and the correlation specified by its owner.
Specifications and service checks
If the rod set is lengthened midway through a field record, retain the extension change and the depth reference in the notes. Blow count and depth describe the assembled apparatus at that stage of the run.
- Range and configuration
- The hammer configuration, drop distance, cone dimensions and depth graduations identify the system. Penetration per blow is a field result.
- Manufacturers and models
- Humboldt H-4219-series equipment includes dual-mass field DCP assemblies. CONTROLS lists a field DCP assembly for penetration testing.
- What we check
- Measure the masses and dimensions of the submitted assembly and verify the depth scale. Identify interchangeable cones and rods separately.
- Reference equipment and accuracy
- Use a traceable balance and length references. Select a profile instrument or optical method where a cone-angle tolerance applies.
- When to repeat the check
- Field wear concentrates at the cone and couplings. Inspect between jobs and reverify after a bent rod or worn tip is replaced.